3 above ambient temperature operation, 4 probe insertion depth, E-m-hg2-s-v2.1 – ROTRONIC Hygrogen2 User Manual

Page 48

Advertising
background image

E-M-HG2-S-V2.1

Rotronic AG

Bassersdorf, Switzerland

Document code

Unit

HygroGen2: Humidity and Temperature Generator

with AutoCal/AutoCal+, Remote Control and

Range Extensions options.

Instruction Manual

Instruction Manual for Software

Version 2.1

Document Type

Page 44 of 75

Document title

Figure 32: Humidity Map – standard and extended ranges

5.1.3 Above Ambient Temperature Operation

At high temperature and high relative humidity conditions, it is necessary to allow the unit to stabilise

to ensure minimal condensation. It is recommended that, for making measurements over 45°C and

80%rh, the HygroGen2 temperature be permitted to equilibrate for 20 minutes before high humidity is

introduced. It is also recommended that, under these conditions, the desired RH level is achieved in

incremental steps, allowing stabilisation at each step. If the unit is operated at high temperature and

high humidity for prolonged periods, it is strongly recommended that the chamber is thoroughly

dehumidified prior to switching the unit off.

5.1.4 Probe Insertion Depth

When adjusting or calibrating probes against a reference, make sure that the insertion depth of each

instrument is the same for each probe. Using the same immersion depth substantially reduces the effect

of any temperature and humidity gradients in the chamber.

The immersion depth of any probe inserted in the HygroGen2 should be within the limits of 100-150mm

(4-6”), measured from the inner side of the door. This will reduce the effect of room temperature via

stem conduction on the probe.

© 2014; Rotronic AG

E-M-HG2-S-V2.1

Advertising