B1: validation – requirements and accessories, E-m-hg2-s-v2.1 – ROTRONIC Hygrogen2 User Manual

Page 64

Advertising
background image

E-M-HG2-S-V2.1

Rotronic AG

Bassersdorf, Switzerland

Document code

Unit

HygroGen2 Humidity and Temperature Generator:

Instruction Manual
Appendix

Instruction Manual for Software

Version 2.1

Document Type

Page 60 of 75

Document title

B1: Validation – requirements and accessories

For an ISO 17025 calibration system, you will need to show evidence of qualifying your own system,

and permanently logging temperature gradients within the chamber.

For this purpose, ROTRONIC has developed a special door with detachable mapping rig:

And a PRT mounting ring that can be fitted permanently into the chamber for continuous monitoring:

We recommend the MBW T12 multi-channel thermometer for performing these tasks, which when

used with the MBW/RHS dew point mirrors provide a calculated RH map of the entire space.

http://www.mbw.ch/product_T12.php

Figure 40: HG2-PRTring

Figure 39: HG2-D888888-Map

© 2014; Rotronic AG

E-M-HG2-S-V2.1