E-m-hg2-s-v2.1 – ROTRONIC Hygrogen2 User Manual

Page 12

Advertising
background image

E-M-HG2-S-V2.1

Rotronic AG

Bassersdorf, Switzerland

Document code

Unit

HygroGen2: Humidity and Temperature Generator

with AutoCal/AutoCal+, Remote Control and

Range Extensions options.

Instruction Manual

Instruction Manual for Software

Version 2.1

Document Type

Page 8 of 75

Document title

To empty the HygroGen2

1. Insert the tube of the dosing syringe, with the plunger fully depressed, 10mm into the water

port and extract any water by drawing out the plunger.

2. Depress the collar on the water port to remove the dosing syringe tube. Wipe up any spilt water

with an absorbent cloth immediately.

3. Distilled water discharged from the HygroGen2 may be stored safely and reused at a later date.

Note: The HygroGen2 should always be emptied if it is being shipped or stored (see

section 5.3

Preparing the Unit for Transit or Storage

). This will prevent ingress of water into the chamber should

the unit be inverted and prevent possible frost damage.

Water Quality

The HygroGen2 should only be filled with distilled water.

As a precaution against any microbial and algae water contamination, HygroGen2 incorporates an Ultra

Violet (UV) sterilisation lamp within the water reservoir. No user intervention is required - HygroGen2

auto-cycles the process.

Under no circumstance should the water reservoir be opened by non-qualified personnel. Direct

exposure to UV lamps can cause burns or blindness.

© 2014; Rotronic AG

E-M-HG2-S-V2.1