B2: HygroGen2 / Screen Messages

On Screen Message:

Fault/Action:

HygroClip NC

HygroClip not connected or faulty. Check and replace as required.

Contact service centre if message persists.

HygroClip error

Temperature or RH values are incorrect within the HygroGen2’s

programming. Remove Probe and check values. Replace if found to be

faulty. Contact service centre if message persists.

HC null data

HygroClip communication; invalid data output. Check HygroClip and

replace. Contact service centre if message persists.

Desiccant probe NC

Desiccant probe connection failure. Service centre repair only. Manual

control of desiccant condition enables use to continue.

Desiccant probe null data

Desiccant probe communication; invalid data output. Service centre

repair only. Manual control of desiccant condition enables work to

continue.

Low Water level

Check water level, empty and refill as required, see

section 2.4

. If

message persists, manual control of water level allows use to continue

until next service

High Water level

Water level is too high; empty completely and refill in accordance with

section 2.4

. Contact service centre if message persists.

Humidification Disabled

Displays when temperature is changing. Wait for temperature to

stabilise.

Desiccant wet

Desiccant condition warning. Replace desiccant (see

section 8.2

).

Heating Disabled

Over-temperature event. Heating shut down for safety reasons.

Contact service centre.

Sample Loop error

Temperature fault (sample loop temperature sensor or heater).

Operation can continue at dew points below ambient temperature.

Water Sensor error

Fault with level sensor. Manual supervision of water level allows use

to continue until next service.

UV error

UV Lamp failure. Contact service centre.

Service Due

More than 3,000 hours in use since last service. Contact your service

centre to arrange for the HygroGen2 to be serviced.