APPENDIX B

SERVICE LOG ERROR MESSAGES

373XXA MM

B-5

Error Message

Description

Corrective Action

ATTENUATOR UNAVAILABLE

Option 6 Port 2 Test Step Attenuator
is not installed.

Install Option 6 Step Attenuator,

DIFFERENT H/W SETUP. RECALL
ABORTED

Model and/or options is (are) different
from the recalled setup.

Reconfigure system to duplicate the
hardware setup that was used to store
the saved data.

DIFFERENT S/W VERSION, RECALL
ABORTED

Saved state not compatible with soft-
ware version or options.

Load compatible software (S/W) ver-
sion and retry.

FREQUENCIES HAVE REACHED
UPPER LIMIT

Frequencies being defined in Multiple
Source mode have reached upper lim-
its of Sources.

Redefine frequencies to not exceed
limits of Sources.

MEMORY LOCATION CORRUPTED

Requested memory location is cor-
rupted.

None. If problem reoccurs after stor-
ing a new setup, contact ANRITSU
Customer Service.

NO BANDS ARE STORED

No frequency bands have been de-
fined and stored.

Define and store frequency bands to
turn on Multiple Source mode.

NO STORED MEMORY DATA

No data is stored in memory for dis-
play or trace math.

Store or re-save measurement data.

OPTION NOT INSTALLED

Selected an option that is not in-
stalled.

None.

OUT OF CAL RANGE

Entered values out of the selected
calibration range.

Change calibration range or re-enter
values that are within the current
range.

OUT OF H/W RANGE

Entered value is out of the instru-
ment’s hardware range.

Re-enter values that are within range.

OUT OF RANGE

Entered value is out of range.

Re-enter values that are within range.

RECEIVER OUT OF RANGE BY
EQUATION

Equation defined in Multiple Source
mode places receiver frequency out of
range when attempting to store band.

Redefine frequency.

SOURCE 1 OUT OF RANGE BY
EQUATION

Equation defined in Multiple Source
mode places Source 1 frequency out
of range when attempting to store
band.

Redefine frequency.

SOURCE 2 OUT OF RANGE BY
EQUATION

Equation defined in Multiple Source
mode places Source 2 frequency out
of range when attempting to store
band.

Redefine frequency.

Table B-1. Operational Error Messages (1 of 2)