Troubleshooting system has error code – Anritsu 373XXA User Manual

Page 67

Advertising
background image

TROUBLESHOOTING

SYSTEM HAS ERROR CODE

373XXA MM

5-13

Item

Description

SWP

Sweep direction:

F = forward (S11, S21), R = reverse (S22, S12)

PWR

System source power setting

SYS

System frequency setting

A

The calibration results for the Test A Channel PCB (A3): This is a
series of 16 values divided into groups of four. Each group should
have values that are approximately

1

4

that of those in the previous

group.

The first group data values should be approximately: 1.00 0.00
0.00 1.00. The values of the next group should be

1

4

that of those

values i.e., 0.25 0.00 0.00 0.25. The same applies for the values
in the remaining two groups.

B

The calibration results for the Test B Channel PCB (A6). See de-
scription for “A” above.

R

The calibration results for the Reference A (or B) Channel PCB
(A4). See description for “A” above.

<OVL>

Indicates the value was too far out of range to be measurable.

PHSLCK

Source Locking: L=Source Lock Mode, T=Tracking Mode, S=Set On
Mode

Table 5-3. Service Log Snap Shot Data Items