A8 ADJUSTMENT
(37397A)

The adjustments are important to the operation of the VNA above 56
GHz. Improperly adjusted A8 Potentiometers may cause lock failures.
Perform these adjustments as follows:

q

The A8 PCB is replaced.

q

44733 Buffer Amp is replaced.

Calibration

Procedure

Perform the following steps:

Equipment Required

None

NOTE

Allow the 373XXA to warm-up at least 30 minutes prior to
performing calibration, and perform the ALC calibration
(paragraph 6-3) before doing this calibration.

Procedure

Step 1.

Connect a V Throughline (3670V50-2 with FF Adapter) be-
tween Ports 1 and 2.

Step 2.

Set the 373XXA to sweep from 50 to 65 GHz.

Step 3.

Set RF power to 0 dB (default value).

Step 4.

Using Diagnostics menu, view Non-Ratio parameters.

Step 5.

View Channel 1 (Ref A) only, and activate Autoscale.

Step 6.

Adjust the “RA” potentiometer on top of A8 for the flattest
trace display. A small discontinuity at 56 GHz is normal.

Step 7.

Perform Step 6 for each of the four Non-Ratio channels.

Post Calibration

Actions

After the calibration process is completed, perform the following ac-
tions, as appropriate:

If calibration passes:

Save the calibration data to (hard) disk, as follows:

Step 1.

Press the

Option Menu

key (Enhancement key group) to dis-

play the OPTIONS menu.

Step 2.

Select DIAGNOSTICS from menu.Then select in se-
quence: H/W CALIBRATIONS |DISK OPERATIONS
|SAVE TO HARD DISK|HW_CAL.SLT.

ADJUSTMENTS

A8 ADJUSTMENT (37397A)

373XXA MM

6-13