Measurement conditions check list 5-23, Troubleshooting measurement in question – Anritsu 373XXA User Manual

Page 77

Advertising
background image

q

Condition of the 373XXA.

q

Quality and condition of the interface connections and connec-
tors.

q

Quality and condition of the calibration components,
through-lines, adapters and fixtures.

q

Surrounding environmental conditions at the time of the mea-
surement.

q

Selection and performance of the calibration for the DUT being
measured.

Measurement

Conditions Check List

When determining possible measurement problems, check the follow-
ing items:

1.

Check the DUT and the calibration conditions:

q

ensure the Calibration Components Coefficients data has been
installed into the system from the Calibration Kit in use.

q

ensure the proper calibration was done for the device being
measured:

n For high insertion loss device measurements the calibra-

tion should include isolation, high number of averages,
and lower IF Bandwidth settings during the calibration.

n For high return loss device measurements the calibra-

tion should also include a sliding load calibration.

q

Check DUT mating connector(s) condition and pin depth.

q

Measure an alternate known good DUT, if possible.

q

Check if the environment is stable enough for the accuracy re-
quired for the DUT measurement.

n The system should not be subjected to variations in tem-

perature.

n The system should not be placed in direct sun light or

next to a changing cooling source, such as a fan or air
conditioning unit.

2.

Check the calibration using known good components from the cal-
ibration kit. If measurements of these devices do not produce good
results:

q

Check through-cable stability including condition and pin
depth. Replace with a known good cable, if necessary.

q

Check condition and pin depth of calibration kit components.
Replace with known good components, if necessary.

q

Check condition and pin depth of test port connectors. Replace
with known good ones if necessary.

TROUBLESHOOTING

MEASUREMENT IN QUESTION

373XXA MM

5-23

Advertising