Signal paths troubleshooting, Figure 5-4 – Anritsu 373XXA User Manual

Page 80

Advertising
background image

5-26

373XXA MM

SIGNAL PATHS

TROUBLESHOOTING

YIG Oscillator

X4

(SQM)

Reverse

Switched Filter

Down

Converter

A21A2 PCB

Sampler/Buffer Amp

X2

(SDM)

3 dB pad

X4

(SQM)

Forward

MUX Coupler

(reverse)

3 dB Pad

Transfer Switch

(Ref A/ Ref B

Control)

37 GHz

HiPass Filter

37 GHz

HiPass Filter

Shaped Pad

Shaped Pad

MUX Coupler

(forward)

A21A1 PCB

A4 PCB

Reference IF Processor

SPDT Switch

J4

J5

Ref A

Ref B 38-65 GHz

Ref A 38-65 GHz

Ref B

A8 PCB

Source Lock/ RF

Component Control

Power Amp

Second LO

(A2 PCB)

(25-272MHz)

First LO

(A1 PCB)

(357-536MHz)

Ref A 40MHz - 38GHz

Ref B 40MHz - 38GHz

Rear

Panel

Access

Loop **

40MHz-65GHz (Ref A)

40MHz -

65GHz (Ref B)

ALC Control PCB

49808-3

Ref A

J7

J8

Ref B

2.5 MHz CW

2.5 MHz CW

J4

J2

J1

J1

J3

Ref A 38-65 GHz

Ref B

DC

voltage

38-65

GHz

DC voltage (6 MHz / Volt)

J4

J1

J4

J4

J5

J5

DCV

DCV

J1

J3

J2

YIG tuning currents

Digital and Analog Control

40MHz-

2GHz

6.3-

8.3GHz

2-20GHz

40MHz-

19GHz

9.5-

16.25GHz

40MHz-38GHz

SDM Control

Bias Voltages to YIG,

Switched Filter

Down Converter, SAM

Control of SPDT

switch, transfer switch,

buffer amp, power

amp, ALC control

J5

J5

J3

J1

3 dB pad

3 dB pad

16.8 GHz

Lowpass

Filter

15 dB

Pad

NOTE: Sampler/Buffer amplifiers are individually matched to

Power Amplifiers. If either fails, both should be replaced with

another matched set, part number ND53249.

3 dB Pad

** Indicates not present on model 37297B or 37277B

3 dB pad

Figure 5-4.

37397A Source Lock Signal Path Diagram

Advertising