Appendix d performance specifications, Performance specifications appendix d, D-4 373xxa mm – Anritsu 373XXA User Manual

Page 183

Advertising
background image

Measurement Time vs. Span for 101 Data Points (typi-
cal):

Frequency Span

Time (ms)

40 MHz to 40 GHz

500

20 GHz to 40 GHz

400

10 GHz to 11 GHz

250

TEST PORT CHARACTERISTICS

The specifications in the following table apply when the
proper Model 34U Universal Test Port Adapters are con-
nected, with or without phase equal insertables, to the test
set ports and calibrated with the appropriate Anritsu or
other designated calibration kit at 23

°

C

±

3

°

C using the

OSL calibration method with a sliding load to achieve
12-Term error correction (A 90 minute warm-up time is rec-
ommended.)

MEASUREMENT CAPABILITIES

Number of Channels:

Four independent measurement

channels.
Parameters:

S11, S21, S22, S12, or user-defined com-

binations of a1, a2, b1, and b2. All measurements are
made without the need to manually reverse the test device.
Measurement Frequency Range:

Frequency range of

measurement can be narrowed within the calibration range
without recalibration. CW mode permits single frequency
measurements, also without recalibration. In addition, the

system accepts N discrete frequency points where
2

£

N

£

1601.

Domains:

Frequency Domain, CW Draw, and optional

High Speed Time (Distance) Domain.
Formats:

Log Magnitude, Phase, Log Magnitude and

Phase, Smith Chart (impedance), Smith Chart (Admit-
tance), Linear Polar, Log Polar, Group Delay, Linear Magni-
tude, Linear Magnitude and Phase, Real, Imaginary, Real
and Imaginary, SWR, Power.
Data Points:

1601 maximum. Data points can be

switched to a value of 801, 401, 201, 101, or 51 points
without recalibration (if 1601 points were used in the cali-
bration). In addition, the system accepts an arbitrary set of
N discrete data points where: 2

£

N

£

1601. CW mode per-

mits selection of a single data point without recalibration.

Reference Delay:

Can be entered in time or in distance

(when the dielectric constant is entered). Automatic refer-
ence delay feature adds the correct electrical length com-
pensation at the push of a button. Software compensation
for the electrical length difference between reference and
test is always accurate and stable since measurement fre-
quencies are always synthesized. In addition, the system
compensates reference phase delay for dispersive trans-
mission media, such as waveguide and microstrip.
Markers:

Six independent markers can be used to read

out measurement data. In delta-reference marker mode,
any one marker can be selected as the reference for the

D-4

373XXA MM

PERFORMANCE SPECIFICATIONS

APPENDIX D

Connector

Frequency

(GHz)

Directivity

(dB)

Source Match

(dB)

Load Match

(dB)

Reflection

Frequency

Tracking (dB)

Transmission

Frequency Track-

ing (dB)

Isolation

(dB)

GPC-7

0.0225

>52

>44

>52

±

0.003

±

0.004

>105

2.0

>52

>44

>52

±

0.003

±

0.004

>115

18

>52

>42

>52

±

0.004

±

0.012

>112

GPC-7 LRL

Calibration

2.0

>60

>60

>60

±

0.001

±

0.001

>115

18

>60

>60

>60

±

0.001

±

0.001

>112

N-Type

0.0225

>46

>36

>46

±

0.004

±

0.004

>105

2.0

>44

>36

>44

±

0.004

±

0.004

>115

18

>40

>32

>40

±

0.005

±

0.012

>112

3.5 mm

0.0225

>44

>40

>44

±

0.005

±

0.030

>105

2.0

>44

>40

>44

±

0.005

±

0.030

>115

20

>44

>38

>44

±

0.006

±

0.050

>110

26.5

>44

>34

>44

±

0.006

±

0.070

>102

K

0.0225

>42

>40

>42

±

0.005

±

0.030

>105

2.0

>42

>40

>42

±

0.005

±

0.050

>115

20

>42

>34

>42

±

0.006

±

0.070

>110

40

>38

>34

>38

±

0.006

±

0.080

>100

V

0.04

>70

>36

>40

±0.050

±0.030

>105

2.0

>40

>36

>40

±0.050

±0.050

>115

20

>40

>36

>40

±0.060

±0.070

>110

40

>36

>32

>36

±0.060

±0.080

>100

50

>34

>30

>34

±0.080

±0.100

>90

65

>34

>28

>34

±0.100

±0.120

>80