3 lo 2 calibration 6-4, Calibration procedure 6-4, Lo 2 calibration adjustments – Anritsu 373XXA User Manual

Page 84

Advertising
background image

If calibration passes:

Save the calibration data to (hard) disk, as follows:

Step 1.

Press the

Option Menu

key (Enhancement key group) to dis-

play the OPTIONS menu.

Step 2.

Select DIAGNOSTICS from menu.Then select in se-
quence: H/W CALIBRATIONS |DISK OPERATIONS
|SAVE TO HARD DISK|HW_CAL.LO1.

NOTE

Also save the calibration data to floppy disk
for archival purposes as a backup in case of
a hard disk failure. Label the disk with cali-
bration type, date, and system model and se-
rial numbers. Set the copy protect tab on the
disk. Properly safeguard the disk so that it
will be available for later use, as needed.

If calibration fails:

Step 1.

Repeat the calibration. If it still fails, then proceed to
Chapter 5, Troubleshooting.

6-3

LO 2 CALIBRATION

This procedure uses the 373XXA internal diagnostics and calibration
menus to adjust the A2 2nd LO PCB assembly. Perform this calibation
procedure if:

q

The A2 PCB is replaced.

q

BBRAM chip on A9 Processor PCB is replaced and the LO2 Cali-
bration data was not previously saved on disk (that is, data is not
available for recall from a floppy disk).

Calibration

Procedure

Perform the following steps:

Equipment Required

None

NOTE

Allow the 373XXA to warm-up at least 30 minutes prior to
performing calibration.

Procedure

Step 1.

Press the

Option Menu

key (Enhancement key group) to dis-

play the OPTIONS menu.

LO 2 CALIBRATION

ADJUSTMENTS

6-4

373XXA MM