System has error code troubleshooting, See sheet 2, Figure 5-2 – Anritsu 373XXA User Manual

Page 68: Go to 1

Advertising
background image

SYSTEM HAS ERROR CODE

TROUBLESHOOTING

5-14

373XXA MM

Fails both

Forward and

Reverse?

Lock Failure DE

Code 6XXX)

(Error

Fails below

38 GHz?

Fails above

38 GHz?

Fails above

38 GHz?

Normal power at

Port 1?

Normal power at

Port 1?

Normal power at

Port 1?

OK 40MHz

to 272MHz?

Power Amp or

LO1 Failure

Buffer Amp, A4

or A8 Failure

A8, Buffer Amp

Failure

SDM, SW filter, Xfer

Switch, or source

PCB failure

Not logical. Call

Customer Service

Buffer Amp, Power

Amp, or A1 Failure

A8,YIG Osc, SW filter,

or Source PCB Failure

SPDT Switch, SW

Filter, A8 or Source

PCB Failure

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

No

No

No

No

No

No

No

Yes

Yes

NOTE: Sampler/Buffer amplifiers are individually matched to

Power Amplifiers. If either fails, both should be replaced with

another matched set, part number ND53249.

Go to

1

See Sheet 2

Figure 5-2.

Error Code DE Lock Failure Troubleshooting (37397A) (1 of 3)

Advertising