Service log error messages appendix b, B-6 373xxa mm – Anritsu 373XXA User Manual

Page 162

Advertising
background image

SERVICE LOG ERROR MESSAGES

APPENDIX B

B-6

373XXA MM

Error Message

Description

Corrective Action

STANDARD CAL NOT VALID FOR
WAVEGUIDE

Cannot use the standard method
when calibrating with waveguide.

Use the Offset Short method with
waveguide.

START F FOLLOWS PREVIOUS
STOP F

Start frequency of current band imme-
diately follows stop frequency of previ-
ous band. Cannot be modified.

None.

START MUST BE LESS THAN STOP

Entered start frequency is greater
than the stop frequency.

Re-enter frequency values such that
the start frequency is lower than the
stop frequency.

STEP IS TOO LARGE

Entered harmonic frequency extends
the stop out of range.

Re-enter so that harmonic frequency
is within range.

STOP IS OVER RANGE

Entered value exceeds the instru-
ment’s stop frequency.

Re-enter stop frequency.

SYSTEM NOT CALIBRATED

373XXA is uncalibrated for the se-
lected measurement values.

Perform a measurement calibration.

TOO FEW POINTS, 2 MINIMUM

Entered too few discrete fill points, 2
is minimum.

Re-enter data points.

TOO MANY POINTS, 1601 MAXI-
MUM

Entered too many discrete fill points,
1601 points are the maximum al-
lowed.

Re-enter data points.

UNDEFINED DIVIDE BY ZERO

Denominator cannot be zero in equa-
tion.

Make denominator a value other than
zero.

WINDOW TOO SMALL

Attempted to set time domain range
smaller than allowed

Re-enter larger time range.

OUT OF WINDOW RANGE

Attempted to set time domain range
larger than allowed

Re-enter values within allowed range.

Table B-1. Operational Error Messages (2 of 2)