Post calibration actions 6-5, 4 frequency calibration 6-5, Adjustments frequency calibration – Anritsu 373XXA User Manual

Page 85

Advertising
background image

Step 2.

Select DIAGNOSTICS from menu.Then select in se-
quence: H/W CALIBRATIONS then LO2 CALIBRA-
TION.

Step 3.

Follow the directions displayed on the 373XXA CRT screen,
until calibration is completed.

Post Calibration

Actions

After the calibration process is completed, perform the following ac-
tions, as appropriate:

If calibration passes:

Save the calibration data to (hard) disk, as follows:

Step 1.

Press the

Option Menu

key (Enhancement key group) to dis-

play the OPTIONS menu.

Step 2.

Select DIAGNOSTICS from menu.Then select in se-
quence: H/W CALIBRATIONS |DISK OPERATIONS
|SAVE TO HARD DISK|HW_CAL.LO2.

NOTE

Also save the calibration data to floppy disk
for archival purposes as a backup in case of
a hard disk failure. Label the disk with cali-
bration type date, and system model and se-
rial numbers. Set the copy protect tab on the
disk. Properly safeguard the disk so that it
will be available for later use, as needed.

If calibration fails:

Step 1.

Repeat the calibration. If it still fails, then proceed to
Chapter 5, Troubleshooting.

6-4

FREQUENCY
CALIBRATION

This procedure uses the internal 373XXA diagnostics and calibration
menus, in conjunction with a suitable frequency counter, to adjust the sig-
nal source frequencies throughout the range of the 373XXA model being
calibrated. Perform this calibration procedure if:

q

The Source FM/Lock Linearity test in Chapter 3, Operational
Test, fails.

q

Other testing or troubleshooting reveals a possible problem with
the signal source frequency accuracy or phase lock loop.

q

Any of the following assemblies are replaced:

n A21A1 Source YIG/Bias
n A21A2 Source Controller
n Down Converter

ADJUSTMENTS

FREQUENCY CALIBRATION

373XXA MM

6-5