System has error code troubleshooting, Figure 5-2 – Anritsu 373XXA User Manual

Page 70

Advertising
background image

SYSTEM HAS ERROR CODE

TROUBLESHOOTING

5-16

373XXA MM

Reverse Fails

Only?

Yes

No

Fails below

38 GHz?

Fails above

38 GHz?

Fails above

38 GHz?

Normal power at

Port 2 (S12)?

Normal power at

Port 2 (S12)?

Normal power at

Port 2 (S12)?

Buffer Amp, A4

or A8 Failure

A8, A4, or Buffer

Amp Failure

A8, transfer sw

failure

Buffer Amp, A8, or

A4 Failure

Ref B MUX Failure

Ref B SQM, SPDT

Switch, A8, or MUX Failure

Yes

Yes

Yes

Yes

Yes

Yes

No

No

No

No

No

No

2

Not logical. Call

Customer Service

Not logical. Call

Customer Service

From Sheet 1

Figure 5-2.

Error Code DE Lock Failure Troubleshooting (37397A) (3 of 3)