APPENDIX B

SERVICE LOG ERROR MESSAGES

373XXA MM

B-9

Error Message

Description

GET received during PM recep-
tion

Generated when the GPIB Command ‘Group Execute Trigger’ is received during the
reception of a program message but before its proper termination with the end mes-
sage. The partial program message up to but not including the ‘Group Execute Trig-
ger’ will be executed. Execution of the Group Execute Trigger and any subsequent
program message elements received before the end message will be skipped.

7205 GPIB EXECUTION ERROR DESCRIPTIONS

Not permitted in a DDT com-
mand sequence

When executing a defined device trigger command sequence, a forbidden command
was detected.

Too much Arbitrary Block data

The arbitrary block supplied contained more data than was necessary for the currently
defined 373XXA state. This can occur when graph types, start/stop frequencies or
data points are changed.

Insufficient Arbitrary Block data

The arbitrary block supplied did not have enough data for the currently defined
373XXA state. This can occur when graph types, start/stop frequencies or data points
are changed.

Invalid parameter for current
graph type

An attempt was made to program a non-existent parameter for the current graph type.
For instance, a Smith chart does not have a reference or reference line position (mne-
monics OFF and REF).

Parameter out of range

An attempt was made to program an out of integer range value for a parameter. This
error is detected by the GPIB MANAGER when converting and rounding to the appro-
priate integral size (signed/unsigned char/short or long).

Parameter value not permitted

A parameter value was not found in the list of permissible values for that parameter.

CW marker sweep not permitted
in time domain

The mnemonics M1C, M2C, M3C, M4C, M5C and M6C are forbidden in time domain.

Parameter unavailable in fre-
quency domain

The mnemonic ODV and OTV are forbidden in frequency domain.

Port 2 Test Attenuator (OPT 6)
not installed

The mnemonic TA2 is forbidden when the attenuator is not installed.

Time Domain (OPT 2) not in-
stalled

An attempt was made to use one of the time domain mnemonics when the option is
not installed.

Return to Local not permitted in
Local Lockout

The mnemonic RTL failed due to being in the Local Lockout mode.

Calibration does not exist

An attempt was made to turn on flat power correction or vector error correction when
the corresponding calibration does not exist.

Cal term not available

An attempt was made to get a calibration term which does not exist for the current
calibration type.

Table B-3. GPIB-Related Error Messages (2 of 8)