TROUBLESHOOTING

SYSTEM HAS ERROR CODE

373XXA MM

5-15

Fails in

Forward only ?

Fails below

38 GHz?

Fails above

38 GHz?

Fails above 38

GHz?

Normal power at

Port 1 (S21)?

Normal power at

Port 1 (S21)?

Normal power at

Port 1 (S21)?

Buffer Amp, A4 or

A8 Failure

A8, A4, or Buffer

Amp Failure

Transfer Switch

or A8 Failure

Buffer Amp, A4,

or A8 Failure

Ref A MUX

Coupler Failure

SPDT Switch, Ref A

SQM, A8 or MUX

Coupler Failure

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

No

No

No

No

No

No

1

Go to

2

Not logical. Call

Customer Service

See Sheet 3

From Sheet 1

Figure 5-2.

Error Code DE Lock Failure Troubleshooting (37397A) (2 of 3)