Isolation procedures 5-19, Troubleshooting source/test set/receiver – Anritsu 373XXA User Manual

Page 73

Advertising
background image

By examining these displays in conjunction with other failure symp-
toms, the failed assembly can be located.

Example: Suppose the Non-Ratio Parameter screen display shows
power holes above 270 MHz on all traces and in both sweep directions.
A possible cause for this condition is a failure of the Power Amplifier
assembly at certain frequencies. (Refer to Analog Subsystem Block Di-
agram in Figure 7-2 on page 7-6).

LO1

This function allows user to verify whether the A1 LO1 PCB is operat-
ing properly.

LO2

This function allows the user to verify whether the A2 LO2 PCB is op-
erating properly.

SOURCE

This function allows the user to verify whether the Source Module is
operating properly.

Isolation Procedures

The procedures in the following paragraphs will help you isolate the fault
to the major assembly.

Check #1

Proceed as follows:

Step 1.

Press

Option Menu

key and select the DIAGNOSTICS

menu option.

Step 2.

Select the READ SERVICE LOG menu option to display
the error entry associated with the system failure. Observe
the snap shot data included as part of the error entry and
note the sweep direction indication for each error code. (Re-
fer to Table 5-3.)

TROUBLESHOOTING

SOURCE/TEST SET/RECEIVER

373XXA MM

5-19

OPTIONS

TRIGGERS

REAR PANEL

OUTPUT

DIAGNOSTICS

MULTIPLE SOURCE

CONTROL

PRESS <ENTER> TO

SELECT

DIAGNOSTICS

START SELF TEST

READ SERVICE LOG

INSTALLED OPTIONS

PERIPHERAL TESTS

TROUBLESHOOTING

(FOR SERVICE

ONLY)

HW CALIBRATIONS

(FOR SERVICE

ONLY)

PRESS <ENTER> TO

SELECT