Chapter 6
Adjustments

6-1

INTRODUCTION

This chapter contains procedures that are used to restore the calibration
of the 373XXA signal source and the related source lock system assem-
blies. Use these procedures after various signal source related assemblies
have been replaced due to troubleshooting or repair activities.

Please insure you have read and fully understand the servicing con-
cepts for the 373XXA presented in Chapter 1 prior to continuing with
this chapter.

6-2

LO1 CALIBRATION

This procedure uses the 373XXA internal diagnostics and calibration
menus to adjust the A1 1st LO PCB assembly. Perform this calibation
procedure if:

q

The A1 PCB is replaced.

q

BBRAM chip on A9 Processor PCB is replaced and the Source
Lock Threshold Calibration data was not previously saved on
disk (that is, data is not available for recall from a floppy disk).

Calibration

Procedure

Perform the following steps:

Equipment Required

None

NOTE

Allow the 373XXA to warm-up at least 30 minutes prior to
performing calibration.

Procedure

Step 1.

Press the

Option Menu

key (Enhancement key group) to dis-

play the OPTIONS menu.

Step 2.

Select DIAGNOSTICS from menu.Then select in se-
quence: H/W CALIBRATIONS then LO1 CALIBRA-
TION.

Step 3.

Follow the directions displayed on the 373XXA CRT screen,
until calibration is completed.

Post Calibration

Actions

After the calibration process is completed, perform the following ac-
tions, as appropriate:

373XXA MM

6-3