Special precautions: 4-4, 3 measurement accuracy 4-4, Verification software 4-4 – Anritsu 373XXA User Manual

Page 50: Measurement accuracy performance verification, 4 373xxa mm

Advertising
background image

Error Correction:

Perform software-guided 12 term calibration.

Special Precautions:

When performing the procedures in this chapter, observe the following
precautions:

q

Minimize vibration and movement of system and attached compo-
nents and through-cable.

q

Clean and check pin depth and condition of all adapters,
through-cable(s), and calibration components.

q

Pre-shape the through-cable(s) so as to minimize its movement dur-
ing calibration and measurement activities.

q

Ensure the system remains motion free in stable environmental
conditions as defined above throughout warm-up, calibration, and
measurement activities.

4-3

MEASUREMENT
ACCURACY

The verification procedures described in the following paragraphs verify
the published measurement accuracy and measurement traceability* for
the 373XXA.

Verification Software

These procedures use the ANRITSU VNA Verification Software, PN:
2300-237, and the NIST traceable impedance transfer standards con-
tained in the appropriate ANRITSU Verification Kit. The Standard Con-
ditions and Special Precautions described in paragraph 4-2 should be
observed when performing these procedures.

The appropriate ANRITSU Verification Kit to be used for performing
these procedures is model dependent, as follows:

q

Models 37317A and below, with Opt 7A: Verification Kit 3667

q

All Models using K Connector test ports: Verification Kit 3668

q

Models 37397A: Verification Kit 3669B

The impedance transfer standards contained in these kits are:

q

20 dB Attenuation Standard

q

50 dB Attenuation Standard

q

50 Ohm Air Line Standard

q

25 Ohm Mismatch (Beatty) Standard

The verification software performs the following functions:

q

It guides the user through a full 12 Term calibration of the 373XXA.

q

It guides the user through the measurement of the S-parameters of
the NIST traceable impedance transfer standards (below).

q

It verifies the measured values are within the specified measure-
ment uncertainty limits.

q

It indicates the pass/fail status of the measurements on the CRT
display. It can also provide a hard copy printout of the measured

MEASUREMENT ACCURACY

PERFORMANCE VERIFICATION

4-4

373XXA MM

Traceability to the U.S. National Institute of Standards and Technology (NIST)