8 a21a2 adjustment (37397a) 6-14, Calibration procedure 6-14 – Anritsu 373XXA User Manual

Page 94

Advertising
background image

NOTE

Also save the calibration data to floppy disk
for archival purposes as a backup in case of
a hard disk failure. Label the disk with cali-
bration type, date, and system model and se-
rial numbers. Set the copy protect tab on the
disk. Properly safeguard the disk so that it
will be available for later use, as needed.

If calibration fails:

q

Repeat the calibration. If it still fails, then proceed to Chapter 5,
Troubleshooting.

6-8

A21A2 ADJUSTMENT
(37397A)

This adjustment ensures the correct RF power level will be available
to drive the source quadrupler modules (SQMs). Perform these adjust-
ments as follows:

q

D454244 is replaced.

q

A21A2 PCB is replaced.

q

C21620 YIG Oscillator is replaced.

NOTE

Failure to perform this adjustment may result in intermit-
tent operation.

Calibration

Procedure

Perform the following steps:

Refer to Table 1-1, located on page 1-9, for further information about
the following equipment:

HP 437B Power Meter with HP 8487A Power Sensor

or:

ANRITSU ML24XXA with Version 2.02 or later, with ANRITSU
Power Sensors MA2474A and SC6230 for 50 and 65 GHz units.

or:

Gigatronics 8541/8542 with 80304A Power Sensor

GPIB cable, WILTRON 2100-2, or equivalent

NOTE

Allow the 373XXA and power meter to warm-up at least 30
minutes prior to performing calibration.

A21A2 ADJUSTMENT (37397A)

ADJUSTMENTS

6-14

373XXA MM