Control limit line buzzer, Control limit line testing, Set limit line y-axis value – Agilent Technologies N9340A User Manual

Page 104

Advertising
background image

6

SCPI Command Reference

102

N9340A User’s Guide

Control Limit line Buzzer

:CALCulate:LLINe[1]:BUZZer[:STATe]
OFF|ON|0|1
:CALCulate:LLINe[1]:BUZZer[:STATe]?

This command toggles the audio warning between
on and off. If data exceeds the limit line the buzzer
will sound an alarm.

Control Limit Line Testing

:CALCulate:LLINe[1]:[STATe] OFF|ON|0|1
:CALCulate:LLINe[1]:[STATe]?

This command toggles the limit line testing
between on and off.

If the limit line is turned on, limit pattern is
automatically turned off.

Set Limit Line Y-axis value

:CALCulate:LLINe[1]:Y <value>
:CALCulate:LLINe[1]:Y?

This command sets limit line Y- axis value.

Limit line Y- axis value is set independently without
affected by the X- axis Units.

*RST State:

Off

Key Access:

[Limit] > {Beep}

*RST State:

Off

Key Access:

[Limit] > {Limit Line}

*RST Value:

0 dBm

Available Unit:

dBm, -dBm, mV, uV

Key Access:

[Limit] > {Limit Line}

Advertising