Figure 8 using the average detector, Figure 8. using the average detector – Agilent Technologies N9340A User Manual

Page 54

Advertising
background image

4

Making Measurements

52

N9340A User’s Guide

Using the Average Detector and

Increased Sweep Time

When the analyzer’s noise masks low- level signals,
changing to the average detector and increasing the
sweep time smooths the noise and improves the
signal’s visibility. Slower sweeps are required to
average more noise variations.

1 Refer to

“Reducing Input Attenuation"

on

page 49,

and follow steps 1, 2 and 3.

2 Press [TRACE] > {More (1 of 2)} > {Detector} >

{Average} to select the average detector.

3 Press [BW/SWP] > {Sweep Time} to set the sweep

time to 500 ms.

Note how the noise smooths out, as there is
more time to average the values for each of the
displayed data points.

4 Press [BW/SWP] > {Avg Type} to change the

average type.

Figure 8 Using the Average Detector