Getting Started

2

N9340A User’s Guide

29

Upgrading Firmware

A N9340A provides an easy and fast access for
firmware upgrade.

1 Setup a folder named N9340DATA in the root

directory of your USB memory stick.

2 Download the firmware update package from

Agilent website (www.agilent.com/find/n9340a)
into the folder named N9340DATA.

3 Insert the USB memory stick into the USB

connector on the top panel of your analyzer.

4 Press [SYS] > {More (2 of 3)} > {Upgrade} >

{Firmware} active the inner updating procedure.
Then the instrument will perform the upgrading
process automatically.

Adding an option

Pressing

[SYS] > {More (2 of 3)} > {Option} > {Add}

brings up a dialog box for entering the option
license code. Use the numeric keypad to input the
option license code and use the

[Enter] key as a

terminator. If the option license code is identified,
an annotation “Option install successfully” will
appear in the status line, else an annotation
“Invalid option licence” will appear in the status
line.

CAU-

CAUTION

Make sure at least 10 minutes consistently power supply

when updating firmware. Power cut off during the

updating process can make damage to the instrument.