Marker to, Mode, Normal – Agilent Technologies N9340A User Manual

Page 86: Freq count

Advertising
background image

5

Key Reference

84

N9340A User’s Guide

Marker To

Access the following marker function menu keys:

• To Center

Sets the center frequency of the analyzer to the
marker frequency. In Delta mode, pressing

[Marker]

> {Marker To} > {To Center} sets the center frequency
to the marker delta value. The function is not
available in zero span.

Key Access:

[MARKER] > {To Center}

• To Ref

Changes the reference level to the active marker
value, moving the marked point to the reference
level (top line of the graticule). In Delta mode,
pressing

[Marker] > {Marker To} > {To Ref} sets the

reference level to the amplitude difference between
the markers.

Key Access:

[MARKER] > {Marker To}

Mode

Accesses the following menu keys that allows you
to change the active marker readout mode.

• Normal

Active the default readout mode.

• Freq Count

Activates a frequency counter mode.

NOTE

When the frequency scale is in log mode, the center

frequency is not at the center of the display.

NOTE

Frequency counter properly only on CW signals or discrete

peaks. For a valid reading, the marker must be >25 dB above

the noise.

Freq Count function is not supported under Marker Delta

condition.

Advertising