Key Reference

5

N9340A User’s Guide

69

Preamp

Toggles the internal preamp between

On and Off.

Preamp results in a correction being applied to
compensate for the gain of the preamp at 20 dBm
so that amplitude readings show the value at the
input connector.

The preamp is set on in frequency bands from

1 MHz to 3 GHz, otherwise the correction is not
applied even though the

PA indication will still

appear.

Key Access:

[AMPTD]

Scale/DIV

Sets the logarithmic units per vertical graticule
division on the display.

Key Access:

[AMPTD]

Scale Type

Toggles the scale type of vertical graticule divisions
between logarithmic and linear.

When scale type sets to

Log, a Logarithmic units is

activated ranging from 1 to 10 dB per division.

When Scale Type sets to

Lin, the vertical divisions

are linearly scaled and the default amplitude units
are volts. The top of the display is set to the
reference- level value and the bottom graticule
becomes zero volts. (Each division of the tragically
is one- tenth of the reference level in volts.)

Key Access:

[AMPTD] > {Scale Type}

NOTE

When the preamp is on, a PA indication appears on the left

side of the display.

NOTE

The Scale/Div function is only available when the Scale Type

key is set to Log.