SCPI Command Reference

6

N9340A User’s Guide

103

Control Limit Pattern Testing

:CALCulate:LLINe[1]:PATTern[:STATe]
OFF|ON|0|1
:CALCulate:LLINe[1]:PATTern[:STATe]?

This command toggles the limit pattern testing
between on and off.

If the limit pattern is turned on, limit line is
automatically turned off.

Set Limit Pattern X-axis Units

:CALCulate:LLINe[1]:PATTern:DOMain
FREQ|TIME
:CALCulate:LLINe[1]:PATtern:DOMain?

This command selects how the limit pattern points
are defined: according to frequency, or according to
the sweep time setting of the spectrum analyzer.
The query command returns a string of FREQ or
TIME.

*RST State:

Off

Key Access:

[Limit] > {Limit Pattern}

*RST State:

Frequency

Range:

FREQ|TIME

Key Access:

[Limit] > {Set pattern} > {X axis unit}