Making Distortion Measurements

This section provides information on measuring
and identifying signal distortion

.

Identifying Analyzer Generated Distortion

High level input signals may cause analyzer
distortion products that could mask the real
distortion measured on the input signal. Use trace
and the RF attenuator to determine which signals,
if any, are internally generated distortion products.
In this example, use a signal from a signal
generator to determine whether the harmonic
distortion products are generated by the analyzer.

1 Input a signal (200 MHz, –10 dBm) to the

analyzer

RF IN connector.

2 Set the analyzer center frequency and span:

Press

[PRESET]. (Factory Preset)

Press

[FREQ] > {Center Freq} > 400 > {MHz}.

Press

[SPAN] > 700 > {MHz}.

The signal produces harmonic distortion products
(spaced 200 MHz from the original 200 MHz signal)
Figure 11 Harmonic Distortion