Agilent Technologies N9340A User Manual

Page 58

Advertising
background image

4

Making Measurements

56

N9340A User’s Guide

3 Change the center frequency to the value of the

first harmonic:

Press

[MARKER] > {Peak Search}.

Press

[MARKER] > {Marker To} > {To Center}.

4 Change the span to 50 MHz and re- center the

signal:

Press

[SPAN] > 50 > {MHz}.

Press

[MARKER] > {Peak Search}.

Press

[MARKER] > {Marker To} > {To Center}.

5 Set the attenuation to 0 dB:

Press

[AMPTD] > {Attenuation} > 0 > {dB}.

Press

[MARKER] > {Marker To} > {To Ref}.

6 To determine whether the harmonic distortion

products are generated by the analyzer, first save
the trace data in trace 2 as follows:

Press

[TRACE] > {Trace (2)}.

Press

[TRACE] > {Clear Write}.

7 Allow trace 2 to update (minimum two sweeps),

then store the data from trace 2 and place a
delta marker on the harmonic of trace 2:

Press

[TRACE] > {View}.

Press

[MARKER] > {Peak Search}.

Press

[Marker] > {Delta}.

Please refer to

Figure 12

, the screen shows the

stored data in trace 2 and the measured data in
trace 1. The Marker indicator reads the difference
in amplitude between the reference and active
markers.

8 Press [AMPTD] > {Attenuation} > 10 > {dB} to

increase the RF attenuation to 10 dB.