Making Measurements

46

N9340A User’s Guide

Resolving Signals of Equal Amplitude

In this example a decrease in resolution bandwidth
is used in combination with a decrease in video
bandwidth to resolve two signals of equal
amplitude with a frequency separation of 100 kHz.

Notice that the final RBW selection to resolve the
signals is the same width as the signal separation
while the VBW is slightly narrower than the RBW.

1 Connect two sources to the analyzer input as

shown below.

Figure 3

Fr eq u en c y

En te r

7

M O D

O n/ O ff

RF

4

1

0

2

9

6

3

O n/ O ff

Am p litu de

FM

U tility

LF O ut

Pr ese t

Lo c a l

AM

I/Q

File

Tr ig ge r

Pu lse

M

·

Sw e ep

8

5

Re m o te

Sta n d b y
O n

N 9310 A RF Sig na l G en er a to r 9 kH z - 3.0 GH z

REV ERSE PW R
4W M AX 30 VD C

LF O U T

RF O U T 50

FUN CTIO N S

Fre qu en c y

En ter

7

M O D

O n /O ff

RF

4

1

0

2

9

6

3

O n / O ff

Am p litu de

FM

U tility

LF O u t

Pr eset

Lo c a l

AM

I/Q

File

Trigg er

Pulse

M

·

Sw ee p

8

5

Re m o te

Sta nd b y
O n

N 9310 A RF Sig n a l G en er a to r 9 kHz - 3. 0 GHz

REV ERSE PW R

4W M AX 30 VD C

LF O U T

RF O UT 50

FU N CTIO N S

Directional

coupler

Signal generator

Signal generator

Setup for obtaining two signals

2 Set one source to 300 MHz. Set the frequency of

the other source to 300.1 MHz. Set both source
amplitudes to –20 dBm.

3 Setup the analyzer to view the signals:

Press

[PRESET].

Press

[FREQ] > {Center Freq} > 300 > {MHz}.

Press

[BW/SWP] > {RBW} > 300 > {kHz}.

Press

[SPAN] > {Span} > 2 > {MHz}.