Initiate subsystem, Continuous or single measurements – Agilent Technologies N9340A User Manual

Page 118

Advertising
background image

6

SCPI Command Reference

116

N9340A User’s Guide

INITiate Subsystem

The INITiate subsystem is used to control the
initiation of the trigger. Refer to the TRIGger
subsystem for related commands.

Continuous or Single Measurements

:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?

Selects whether the trigger system is continuously
initiated or not.

This command affects sweep if not in a
measurement, and affects trigger when in a
measurement. A

“measurement” refers to any of

the functions under the

MEAS key. This corresponds

to continuous sweep or single sweep operation
when not in a measurement, and continuous
measurement or single measurement operation
when in a measurement.

When not in a measurement, this command does
the following:

• When

ON at the completion of each sweep cycle,

the sweep system immediately initiates another
sweep cycle.

• When

OFF, the sweep system remains in an “idle”

state until CONTinuous is set to

ON or an

:

INITiate[:IMMediate] command is received.

On receiving the

:INITiate[:IMMediate]

command, it will go through a single sweep
cycle, and then return to the “idle” state.

• The query returns 1 or 0 into the output buffer.

1 is returned when there is continuous sweeping.
0 is returned when there is only a single sweep.

When in a measurement, this command does the
following: