Error Messages

7

N9340A User’s Guide

145

147

Incorrect
alignment data
in EEPROM

Indicates invalid alignment data
occurred.

143

Final IF
overload

Indicates you need to either
increase the input attenuation or
decrease the input signal level.

138

USB device NOT
ready

Indicates the USB device is not
detected.

136

RBW limit to 30
kHz when in fast
sweep

Indicates the analyzer
automatically couple the RBW to
30 kHz when in fast sweep
mode.

135

Error 0xB

Indicates the analyzer encounters
a DLL file open failure. If
analyzer froze, try to restart the
analyzer.

134

Error 0xA

Indicates the analyzer failed at a
dynamic loading process. If
analyzer froze, try to restart the
analyzer.

132

Error 0x8

Indicates some important files
missing.

130

Meas uncal

Indicates the measurement is
uncalibrated due to fast sweeping
through a narrow RBW filter.
Check the sweep time, span, and
bandwidth settings, or use auto
coupling.

119

RF Board Changed

Indicates the RF board was
changed, the analyzer needs
re-load the alignment data.

116

Cannot
communicate with
RF

Indicates the MCU cannot find
the RF board.

112

Error 0x3, DSP
will restart

Indicates an undefined error
occurred when MCU tried to
communicate with DSP. Wait the
analyzer to reboot.

Error Code Error Message

Description