Delta – Agilent Technologies N9340A User Manual

Page 84

Advertising
background image

5

Key Reference

82

N9340A User’s Guide

Delta

Activates a second marker at the position of the
first marker. (If no marker is present, two markers
appear at the center of the display). The amplitude
and frequency (or time) of the first marker is
fixed. The marker number is indicated above the
delta marker, and the same number is indicated
with an R (for example, 1R) above the reference
marker. Use the data controls to position the delta
marker. Annotation in the active function block
and in the upper- right corner of the display
indicates the frequency (or time) and amplitude
differences between the two markers. The markers
will be turned off if the scale type is changed
between log and linear.

Key Access:

[MARKER]

Off

Turns off the active function markers.

Key Access:

[MARKER]

NOTE

Press Delta again moves the reference marker to the active

marker position, so you are allowed to make delta

measurements from differing reference points without having

to turn off the makers and begin again.

NOTE

The delta marker function permits signal-to-noise

measurements provided the signal is a single spectral

component (sinusoid). Place a normal marker on the signal,

press Delta, place the delta marker in the noise, and activate

Marker Noise. The indicated amplitude differences

signal-to-noise/Hz.