Error Messages

144

N9340A User’s Guide

162

Cannot open fast
sweep in this
case

Indicates current settings do not
allow you to open fast sweep.
For example, currently in FFT,
zero span, or any measurement
in power suite does not allow
fast sweep.

159

Error 0x11, DSP
will restart

Indicates DDC failed. If analyzer
froze, try to restart the analyzer.

158

Error 0x10, DSP
will restart

Indicates an internal error
occured, wait the analyzer
to reboot.

If analyzer froze, try

to restart the analyzer.

157

Error 0xF

Indicates an error in internal
communication.

156

Incorrect
alignment file
on flash

Indicates an invalid alignment
file.

153

RF EEPROM
operate fail

Indicates an error when writing
alignment file.

152

Alignment file
oversize

Indicates the alignment file has
error or invalid data.

151

DSP boot fail

Indicates the analyzer
cannot process
measurement,wait the
analyzer to reboot.

If

analyzer froze, try to restart the
analyzer.

150

Mixer overload

Indicates the first mixer in
danger. Either increase input
attenuation or decrease the input
signal level.

149

Mixer saturate

Indicates you need to either
increase the input attenuation or
decrease the input signal level.

148

Error 0xE

Indicates the analyzer reads
unreasonable parameter value. If
analyzer froze, try to restart the
analyzer.

Error Code Error Message

Description