Take new data acquisitions – Agilent Technologies N9340A User Manual

Page 119

Advertising
background image

SCPI Command Reference

6

N9340A User’s Guide

117

• When

ON at the completion of each trigger cycle,

the trigger system immediately initiates another
trigger cycle.

• When

OFF, the trigger system remains in an

“idle” state until CONTinuous is set to

ON or an

:INITiate[:IMMediate] command is received.
On receiving the

:INITiate[:IMMediate]

command, it will go through a single trigger
cycle, and then return to the “idle” state.

• The query returns 1 or 0 into the output buffer.

1 is returned when in a continuous measurement
state. 0 is returned when there is only a single
measurement.

Take New Data Acquisitions

:INITiate[:IMMediate]

This command initiates a sweep if not in a
measurement. If in a measurement, it triggers the
measurement. A “measurement” refers to any
function under the

MEAS key.

Note that the function is enable only when the
continuous is

OFF.

*RST State:

Continuous, or On

Key Access:

[BW/SWP] > {Sweep}

Key Access:

[BW/SWP] > {Single Sweep}