Log Pwr

Averages the data as appropriate for the
logarithmic scaled y- axis.

When average type is set to

Log Pwr, “LgAv”

appears on the left side of the display.

Key Access:

[BW/SWP] > {Avg Type}

Pwr

Performs by converting the trace data from
logarithmic to linear power units, and then
averaging the power trace data.

When average type is set to

Pwr, “PAvg” appears

on the left side of the display.

Key Access:

[BW/SWP] > {Avg Type}

Voltage

Averages the data as appropriate for the linearly
scaled y- axis scale.

When average type is set to

Voltage, “VAvg” appears

on the left side of the display.

Key Access:

[BW/SWP] > {Avg Type}

There are actually four types of averaging
processes within a N9340A. All, except VBW
filtering, are affected by this setting. They are
listed as follows:
• Trace averaging

Averages signal amplitudes on a trace- to- trace
basis. (Press

Log Pwr or Pwr)

• Average detector

Averages signal amplitudes during the time or
frequency interval represented by a particular
measurement point.
• Noise marker

Averages signal amplitudes across measurement
points to reduce variations for noisy signals.
• VBW filtering

Performs a form for averaging the video signal.