One Button Measurement

N9340A provides one- button measurement
functionality on OBW

(Occupied Band Width)

, channel

power and ACPR

(Adjacent Channel Power Ratio)

and

SEM

(Spectrum Emission Mask)

as an easy- to- use

access for your measurement of interest.

Measuring OBW

Occupied Bandwidth (OBW) integrates the power
of the spectrum in the displayed green frame. The
measurement defaults to 99% of the occupied
bandwidth power.
The occupied bandwidth measurement can be made
in single or continuous sweep mode.

Selecting a Method

Press

[MEAS] > {OBW} > {Method} to toggle the

occupied bandwidth measurement mode between

percentage (%)and dBc.
• Setting Percentage (%)

Press

{OBW} > {Method} > {Percent} to edit the

percentage of signal power used when determining
the occupied bandwidth. Press

{%} to set the

percentage ranging from 10.00% to 99.99%.

• Setting dBc

Press

{OBW} > {Method} > {dBc} to specify the power

level used to determine the emission bandwidth as
the number of dB down from the highest signal
peak, within the occupied bandwidth span. This
function is an independent calculation from the
OBW calculation. The x dB Bandwidth result is
also called the emissions bandwidth (EBW). This
will set green frame

(f1 and f2 are the two cross points of the signal
and the pair of vertical lines of the displayed green
frame) to your specified dB value below the
maximum power and compute the total power in
the frame. The emission bandwidth is then
calculated ranging from 0.1 dB to 100 dB (the
default value is set to 26 dB).