Use the arrow buttons to set the tolerance value

Confirm the entered tolerance by pressing

button,

The balance software automatically enters the mode <DOSING> and

displays the target mass value with a minus sign, and in the bottom

line standard mass value – target mass (if such option for the function

<INFORMATION> was selected).

The sign <Min> is shown at the top of the display. It indicates the status of

the mass on the pan with relations to the target mass, below the value

<TARGET MASS – TOLERANCE>

<Min>: mass lower than the

Target Value - Tolerance

<Ok>: mass contained within

tolerance

Target Value +/- Tolerance

<Max>: mass higher than the

Target Value + Tolerance