and then:

1 – 3

4

5

6

7

8 – 16

17

18 - 20

21

22

C

om

m

and

sp

ace

st

ab

ilit

y

m

ar

ker

sp

ace

char

ac

ter

M

as

s

sp

ac

e

uni

t

CR

LF

command

- 1 ÷ 3 - characters

stability marker - [space] – when stable

- [?] – when unstable

- [^] – if there is an error exceeding the maximal range +

- [v] – if there is an error exceeding the maximal range -

characters

- [space] – for positive or [-] values

mass

- 9 characters – aligned to the right

unit

- 3 characters – aligned to the left.

22 COMMUNICATION PROTOCOL

General information

A. A character based communication protocol balance-terminal is designed for

establishing communication between a RADWAG balance and peripheral

devices via RS 232 interface.

B. It consists of commands sent from a peripheral device to the balance and

responses from the balance.

C. Responses are sent from the balance on each receipt of a command as a

reaction to a specific command.

D. Commands forming the communication protocol enable obtaining data on

balance’s status and influencing its operation, e.g.: request measurement

results from the balance, zero indication, etc.