RADWAG PS 510.R2.CT Carat Balance User Manual

Page 165

Advertising
background image

- 165 -

Switch on continuous transmission in current measuring unit

Format: CU1 CR LF

Accessible responses:

CU1_I CR LF

- command understood but not accessible at this moment

CU1_A CR LF

MASS FRAME

- command understood and in progress

- response is mass value in current measuring unit

Responce format

1 2 3 4

5

6

7-15 16

17 18 19 20 21

S U I stability

marker space character mass space unit

CR LF

Switch off continuous transmission in current weighing unit

Format: CU0 CR LF

Accessible responses:

CU0_I CR LF

- command understood but not accessible at this moment

CU0_A CR LF

- command understood and carried out

Set low checkweighing threshold

Format: DH_XXXXX CR LF, where: _ - space, XXXXX – mass format

Accessible responses:

DH_OK CR LF

- command understood

ES CR LF

- command not recognised (mass format incorrect)

Set high checkweighing threshold

Format: UH_XXXXX CR LF, where: _ - space, XXXXX – mass format

Accessible responses:

UH_OK CR LF

- command understood

ES CR LF

- command not recognised (mass format incorrect)

Give value of low checkweighing threshold

Format: ODH CR LF

Accessible response: DH_MASS CR LF – command carried out

Responce format

1

2

3

4-12

13

14

15

16

17

18

19

D H space

mass

space

unit

space

CR

LF

Mass

- 9 characters with right justification

Unit

- 3 characters with left justification

Give value of high checkweighing threshold

Format: OUH CR LF

Accessible response: UH_MASS CR LF – command carried out