12 pipettes calibration, Additional settings of pipettes calibration mode, Pipettes calibration – RADWAG PS 510.R2.CT Carat Balance User Manual

Page 136

Advertising
background image

- 136 -

18.12 Pipettes calibration

(only XA and AS balances)

Pipettes calibration function applies to fixed volume pipettes and adjustable volume

pipettes. During volume testing procedure, the software determines accuracy and

repeatability errors for tested volume. In case of pipettes with adjustable volume

errors for Max, Min and ½ Max volume are estimated.

All pipettes are tested for accordance with requirements of PN-EN ISO 8655:2003.

During the tests dosing repeatability and accuracy are monitored.


In order to ensure the highest accuracy of pipettes calibration process,
maintain the following ambient conditions at a workstation:

Ambient temperature of a pipette, tips and liquid should be kept between

20°C - 25°C with change rate during testing within ± 0.5°C - 111 -

Relative humidity 50 - 75%

and

Use distilled water for pipettes calibration processes

A pipette, tips and distilled water should be stabilized for temperature in

the weighing room. The reference norm advises that minimum

acclimatization time for above mentioned is 2 hours.

18.12.1. Additional settings of pipettes calibration mode

Additional settings have been designed for calibration mode operation. These

settings complete the standard ones.

The options are as follows:

VOLUME DETERMINATION – allows to predefine numer of tested

volumes for a particular pipette. For pipettes with fixed volume please set

parameter <1>, for pipettes with adjustable volume set parameter <2> and

<3>.

MEASUR. NO – allows to predefine numer of measurements for each

tested volume. Number of measurements ranges from 6 to 20.

AUTOMATING TARRING - allows to turn on function of automatic taring

of dosed portion of water after confirmation of measurement (value set to

<YES>)

Remember to select correct values of the above options before pipettees

calibration procedure is performed. The settings should reflect expectations

and needs being a result of working environment.