Returning to the weighing mode, 6 animal weighing, Additional setting for animal weighing mode – RADWAG PS 510.R2.CT Carat Balance User Manual

Page 109: Animal weighing

Advertising
background image

- 109 -

18.5.3. Returning to the weighing mode

18.6

Animal weighing

Working mode <Animal weighing> enables reliable determining mass of

objects in motion. In principle, these types of objects generate unstable

measurement, thus it requires using a different filtering method of

measurement signal.

18.6.1. Additional setting for animal weighing mode

Apart from the standard settings for this mode (described in the weighing

mode), additional settings have been introduced which describe the operation

of the mode.

Available options:

AVERAGING TIME – Amount of time during which measurement records

of weighed sample are analysed. Obtained data is used to determine

measurement result

THRESHOLD - It is a value expressed in mass measuring units. In order

to start measurement, the indication has to exceed value set in the

threshold.

AUTOSTART - Determines start criterion for measurements: whether they

are initiated manually / on pressing a button or selecting START/ or

automatically.

When set on <YES> a measurement starts automatically if indication on

balance’s display exceeds the value of set threshold. The following

measurement can start on unloading the weighed object from balance’s

pan (the indication has to return below the value set in threshold) loading

a new object on the pan and exceeding the threshold’s value by the

indication.

It should be remembered, however, that before the start of the animal weighing

the above-mentioned options should be set on the appropriate values to meet

the expectations and the needs of the working conditions.