RADWAG PS 510.R2.CT Carat Balance User Manual

Page 164

Advertising
background image

- 164 -

Responce format:

1 2 3

4

5

6

7-15 16

17 18 19 20 21

S U space stability

marker space character mass space unit

CR LF

Example

:

S U CR LF

– command sent from a computer

S U _ A CR LF

– command understood and in progress

S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF

– command carried out

response in mass value in current measuring unit.


Immediately send measurement result in current measuring unit

Format: SUI CR LF

Accessible answers:

SUI_I CR LF

- command understood but not accessible at this moment

MASS FRAME

- mass value in current measuring unit is returned immediately

Response format

1 2 3 4

5

6

7-15 16

17 18 19 20 21

S U I stability

marker space character mass space unit

CR LF

Example

:

S U I CR LF

– command from a computer

S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF

– command carried out,

immediate response of mass value in current measuring unit

where: _ - space


Switch on continuous transmission in basic measuring unit

Format: C1 CR LF

Accessible answers:

C1_I CR LF

- command understood but not accessible at this moment

C1_A CR LF

MASS FRAME

- command understood and in progress

- response is mass value in basic measuring unit

Response format

1 2 3

4

5

6

7-15 16

17 18 19 20 21

S I space stability

marker space character mass space unit

CR LF


Switch off continuous transmission in basic measuring unit

Format: C0 CR LF

Accessible responses:

C0_I CR LF

- command understood but not accessible at this moment

C0_A CR LF

- command understood and carried out

Advertising