RADWAG PS 510.R2.CT Carat Balance User Manual

Page 57

Advertising
background image

- 57 -

Example of a report:
.....................................................
--------- Autotest GLP: Report ---------
Balance type PS 3000.R2
Balance ID 400010
User Admin
Software rev. v.0.4.9
Date 2013.07.16
Time 09:17:16
----------------------------------------
Number of measurements 10
Reading unit 0.001/0.01 g
Internal weight mass 1402.094 g
Filter Normal
Value release Fast &Reliable
----------------------------------------
Deviation for Max. -0.118 g
Repeatability 0.0088 g
Signature

........................................

PROCEDURE:

Parameter <P6.9 GLP AUTOTEST > is initiated by pressing

button.

The balance software starts the autotest GLP procedure and it is carried out

automatically from now on to the end.

A user can stop the procedure at any time by pressing

button in the window.

Once the procedure is finished, the value of standard deviation of all

measurements is shown in the main display, and message <RESULT> in the

bottom line, and the final report is printed out automatically (as shown above). It is

possible to print out the report again by pressing

button.

Report is not saved, therefore when abandoning this level, it is deleted from the

balance volatile memory.

To abandon the window press

button and the program returns to the main

menu. To return to weighing press

button repeatedly.