RADWAG PS 510.R2.CT Carat Balance User Manual

Page 167

Advertising
background image

- 167 -


Value release

Format: SS CR LF

Accessible responses:

SS_OK CR LF

- command understood, execution in progress

Command’s function is similar to function of button PRINT located on an overlay,

Internal calibration

Format: IC CR LF

Accessible responses:

IC_A CR LF

IC_D CR LF

- command understood, execution in progress

- calibration completed

IC_A CR LF

IC_E CR LF

- command understood, execution in progress

- range exceeded, time limit for awaiting for a stable result

IC_I CR LF

- command understood but not accessible at this moment

Disable automatic internal calibration of the balance

Format: IC1 CR LF

Accessible responses:

IC1_I CR LF

- command understood but not accessible at this moment

IC1_E CR LF

- operation disabled, e.g. for verified balance

IC1_OK CR LF

- command carried out

Operation is disabled for verified balances.

For non-verified balances the command inhibits internal calibration until it is enabled

via IC0 command or until the balance is turned off. The command does not modifiy

setings specifying calibration start.

Enable automatic internal calibration of the balance

Format: IC0 CR LF

Accessible responses:

IC0_I CR LF

- command understood but not accessible at this moment

IC0_OK CR LF

- command carried out

Operation is disabled for verified balances.

Lock balance keypad

Format: K1 CR LF

Accessible responses:

K1_I CR LF

- command understood but not accessible at this moment

K1_OK CR LF

- command carried out

Command locks the balance keypad (proximity sensors, touch panel) until the moment

of turning the balance off or until sending K0 command.

Unlock balance keypad

Format: K0 CR LF