Printing mode – RADWAG PS 510.R2.CT Carat Balance User Manual

Page 44

Advertising
background image

- 44 -

After measurement stabilization automatic taring of the packaging

mass will proceed (Net marker will appear in the upper part of the

display),

Put product that is to be packed on the packaging;

Display will show a net mass of the product;

Take off the product together with the packaging;

The balance will cancel tara value (the packaging mass saved in

balance storage during the first step of the operating process) after

the gross mass value (set in < P 2.2.3.2 AUTO THRES>

parameter) has been exceeded;

Put packaging of a next product on weighing pan, after

measurement stabilization automatic taring of the packaging mass

will proceed (Net marker will appear in the upper part of the

display);

Put a next product that is to be packed.

For a correct operation of the balance with AUTOTARE function, the threshold

value must be adjusted.

<P 2.2.3.2 AUTO THRES >

parameter is connected with the following functions:

-

automatic tare,

-

automatic operation,

There will be no automatic taring until the balance operates within the gross mass

value range set in < P 2.2.3.2 AUTO THRES > parameter.

10.6. Printing mode

This function enables setting the printing mode, i.e. button activity

.

A user has the possibility of selecting the following settings:

<WHEN STAB> the stable measurement result, along with the settings

for parameter < GLP PRINTOUT> are sent to the printer port. Pressing

the button

, when the result is not stable (no

sign on

display), the balance software will send the measurement result to the

port after reaching the stable condition of the measurement.

<ANYONE> every single pressing of

button results in sending

the measurement result to the printer port along with the settings for

parameter <GLP PRINTOUT>. Every measurement result is sent

(stable and unstable). The <?> sign appears at the beginning of the

Advertising