RADWAG PS 510.R2.CT Carat Balance User Manual

Page 140

Advertising
background image

- 140 -

By means of pipette dose the first portion of water, wait until the

measurement is stable and press ENTER to confirm.

The computer software records the measuremnt. If P2.3.7

AUTOMATING TARRING

parameter is set to value <YES> than the

indication is automatically tared. If P2.3.7 AUTOMATING TARRING

parameter is set to value <NO> than the user must press TARE

button in order to tare the indication before (s)he doses another

porcion of water.

Summary of series of measurements is displayed upon its

completion.

Press ENTER to continue measuring process, press Esc. to stop it.

Upon pressing Esc. button the computer software returns to the main

window.

Upon pressing ENTER, while displaying the summary result for the

last adjustable pipette volume, a report is generated. The report is

printed on a printer connected to the balance (ambient conditions

visible on a report are those that user entered at the beginning of the

calibration procedure). For fixed volume pipette the situation is similar

if the user presses ENTER while displatying summary for volume V1.

The computer software returns to the main window.

Now the user may either start new procedure for the same pipette or

enter new data for a different pipette.