RADWAG PS 510.R2.CT Carat Balance User Manual

Page 139

Advertising
background image

- 139 -

Set the air pressure, next press ENTER for confirmation. The

computer software proceeds to the next step, i.e. entering the first

control volume (V1) for a tested pipette.

Enter (V1) control volume. For fixed volume pipettes (P2.13.5

VOLUME DETERMINATION

parameter set to value <1>) this is the

one and only value to be entered. The computer software proceeds to

the next step, i.e. entering the second control volume (V2) for a tested

pipette.

Enter (V2) control volume. The computer software proceeds to the

next step, i.e. entering the third control volume (V3) for a tested

pipette.

Enter (V3) control volume.

Upon entering the data the computer software proceeds to pipettes

calibration.

Follow the displayed description in order to complete the procedure.