RADWAG PS 510.R2.CT Carat Balance User Manual
Page 169
- 169 -
Example 2
:
Command
:
OMI <CR><LF>
- give accessible working modes
Response
:
OMI <CR><LF>
- accessible working modes are given;
2<CR><LF>
number
4 <CR><LF>
12<CR><LF>
OK <CR><LF>
- command carried out
OMS - Set working mode
Command overview:
Command sets acessible working modes.
Format: OMS_n <CR><LF>
Accessible responses:
OMS_OK <CR><LF>
- command carried out
OMS_E <CR><LF>
- error in-course of command execution, no parameter
or incorrect format
OMS_I <CR><LF>
- command understood but not accessible at this
moment
n
– parameter, decimal value determining working mode number. To see detailed
description go to OMI command.
Example
:
Command
:
OMS_13<CR><LF>
- set working mode Statistics
Response
:
OMS_OK<CR><LF>
- working mode Statistics selected
OMG - Give current working mode
Command overview:
Command gives selected working modes.
Format: OMG <CR><LF>
Accessible responses:
OMG_n_OK <CR><LF>
- command carried out, current working mode given
OMG_I <CR><LF>
- command understood but not accessible at this
moment
n
– parameter, decimal value determining working mode number. To see detailed
description go to OMI command.