RADWAG PS 510.R2.CT Carat Balance User Manual

Page 163

Advertising
background image

- 163 -

Send stable measurement result in basic measuring unit

Format: S CR LF

Accessible responses:

S_A CR LF

S_E CR LF

- command understood and in progress

- time limit exceeded while waiting for stable measurement

result

S_I CR LF

- command understood but not accessible at this moment

S_A CR LF

MASS FRAME

- command understood and in progress

- response is mass value in basic measuring unit

Response format:

1 2-3

4

5

6

7-15 16

17 18 19 20 21

S space stability

marker space character mass space unit

CR LF

Example

:

S CR LF

– command sent from a computer

S _ A CR LF

– command understood and in progress

S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF

– command carried out,

response in mass value in basic measuring unit.

Immediately send measurement result in basic measuring unit

Format: SI CR LF

Accessible responses:

SI_I CR LF

- command understood but not accessible at this moment

MASS FRAME

- response is immediate with mass value in basic weighing

unit

Response format:

1 2 3

4

5

6

7-15 16

17 18 19 20 21

S I space stability

marker space character mass space unit

CR LF

Example

:

S I CR LF

– command sent from a computer

S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF

– command carried out,

immediate response of mass value in basic weighing unit.

Send stable measurement result in current weighing unit

Format: SU CR LF

Accessible responses:

SU_A CR LF

SU_E CR LF

- command understood and in progress

- time limit exceeded while waiting for stable measurement result

SU_I CR LF

- command understood but not accessible at this moment

SU_A CR LF

MASS FRAME

- command understood and in progress

- response is mass value in current measuring unit

Advertising